Product Description
Monitor Patient VS nCompass
100-240V ECG/NIBP/SPO2/Temp
Criticare 81H000XD
The nCompass™ 8100H Series monitor is the perfect choice for busy departments demanding accurate, reliable performance.
The 8100H Series monitor can monitor adult and pediatric patients in virtually every department. The innovative user interface features simplified menus and dedicated function keys for timely setups. The big, bright display with large numerics can easily be seen from across the room.
The 8100H Series monitor mounts quickly and easily to walls and rollstands.
Features and Benefits
Easy-to-read 12.1" active color TFT display.
Easy to learn, easy to use with simplified menus and dedicated function keys.
Mini-Trends can display condensed tabular vital signs data to show recent monitoring activities.
Cardio-Respirogram Trend Display.
Optional invasive cardiac output interface capability.
Compatible with Criticare's 8500 Series Agent Gas Analyzer for anesthetic monitoring.
Technical Specifications
ECG
Number of Leads: 3- or 5-Lead
Lead View: User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead); I, II, or III (3-Lead)
Gain Selection: x0.5, x1, x2, and x4
Frequency Response: Diagnostic: 0.05 to 100 Hz (-3dB);
Monitor: 0.5 to 40 Hz (-3dB);
ST: 0.05 to 40 Hz (-3dB)
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
Pacer Detection/Rejection: Yes
ST Accuracy: ±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm
ST Range: -6.0 to 6.0 mm
NIBP (Non-Invasive Blood Pressure)
Technique: Oscillometric measure upon inflation
Measurement Cycle: < 40 sec. typical
Auto. Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
Max. Allowable Cuff Pressure: Adult/Pediatric: 300 mmHg; Neonate: 150 mmHg
Range: Adult/Pediatric: 30-300 mmHg; Neonate: 20-150 mmHg
Resolution: 1 mmHg
Transducer Accuracy: ±2 mmHg or ±2% or reading, whichever is greater,
over full range
STAT Mode: 5 min of consecutive readings
Clinical Accuracy: Complies with AAMI SP-10:2002
Pulse Oximetry
Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%;
< 50%: unspecified
Method: Dual wavelength LED
Heart (Pulse) Rate
Source: User selectable: Smart, ECG, Pleth, NIBP, IBP
Range: 20 to 300 bpm (ECG, Pleth);
30-240 bpm (NIBP); 30-300 bpm (IBP)
Accuracy: ±1 bpm or ±1% (ECG) whichever is greater
Temperature
Channels: 2
Range, Accuracy: 68° to 113°F (20° to 45°C): ±0.1°C
Display Resolution: ±0.1°
Probe: YSI-400 and YSI-700 Series
Respiration
Rate: 6 to 150 bpm (ECG); 2 to 120 bpm (CO2)
Resolution: 1 bpm
Accuracy: ±2 bpm or ±2%, whichever is greater
Source: User selectable: Smart, CO2, ECG
CO2
Units: mmHg; %; kPa; Torr.
Method: Non-dispersive infrared; autocalibrating.
N2O Compensation: Manual (On/Off).
Range: 0–99 mmHg; 0–12.5%; 0–12.5 kPa; 0–99Torr.
Resolution: 1 mmHg; 0.1%; 0.1 kPa; 1 Torr.
Flow, Sample Rate: 200 ml/min.
Accuracy: ±2 mmHg; ±0.3 vol. %; ±0.3 kPa; ±2 Torr or ±4% of reading, whichever is greater, at 200 ml/min & respiratory rate of ≤ 120 bpm.
Rise Time: 170 ms (10–90%) at 200 ml/min.
IBP (Invasive Blood Pressure)
Channels: 1 or 3.
Transducer Sites: Selectable: ART, PA, LV, RV, LA, RA, CVP, ICP.
Pressure Range: -10 mmHg to 300 mmHg.
Transducer: Freq. Response 0–20 Hz; Impedance >300 ohms;
Sensitivity 5μV/V/mmHg.
Accuracy: ±1 mmHg or ±1%, whichever is greater.
Alarms
Indication: Audible; Visual (EN 60601-1-8)
Silence: Yes; 2 minutes or permanent.
Trends
Memory: 24 hours at 30 second intervals.
Display: Tabular, Graphical.
Display
Screen: 12.1” active color touch screen
Waveforms: 6, maximum
Waveform Types: ECG Leads I, II, III, aVR, aVL, aVF, V; CO2; Respiration; Pleth; IBP; breath-by-breath CO2; ST Histograms; Mini-Trends; OXY-CRG
Sweep Speed: 6.25, 12.5, 25, or 50 mm/sec
Controls
Keys: 10; membrane-activated.
Rotary Knob: Push and rotate; 24 steps/turn.
System Outputs
COM ports: Two RS232-compatible DB9 ports; USB (for software downloads only).
Nurse Call: Contact switch; 3.5 mm stereo phone jack, 14 mm.
Defib/Sync BNC connector.
Video Port: Serial VGA compatible.
DC Out: 5 VDC output for optional wireless accessory.
External Module Port: 8-PIN circular connector.
Printer (optional)
Type: Thermal line printer.
Paper Speed: 12.5, 25 or 50 mm/sec.
Physical Properties
Weight: 12 lbs; 5.4 kg.
Size: 12” (H) x 14.5” (W) x 11.5” (D);
30.5 cm (H) x 36.8 cm (W) x 29.2 cm (D)
Power Requirements
Voltage: 100 to 240 VAC; 50/60 Hz.
Power Consumption: 35 W, typical.
Battery: 1 sealed lead acid.
Battery Life: 2 hours, typical (3 hours, typical for base unit).
Recharge Time: 4 to 6 hours, typical.
Environmental
Operating Temperature: 32° to 104°F (0° to 40°C).
Storage Temperature: 23° to 122°F (-5° to 50°C).
Operating Humidity: 15% to 90%, noncondensing.
Operating Altitude: -300 to 3,000 meters.
Classification
Medical Device: Class II Equipment (IIb EU).
Electrical Protection: Class I Equipment.
Degree of Protection: Type CF, Defibrillator-proof.
Protection Against Ingress: IPX1.
Safety: EN 60601-1.
Custom Configurations
Standard: ECG, Respiration, NIBP, HR, SPO2, Temperature, Arrhythmia and ST analysis, and Nurse Call interface connection.
Optional: CO2, 1 or 3 channel IBP, printer.