Monitor Patient VS nCompass
100-240V ECG/NIBP/SPO2/Temp
Criticare 81H000XD

 Monitor Patient VS nCompass <br>100-240V ECG/NIBP/SPO2/Temp <br>Criticare 81H000XD
Monitor Patient VS nCompass
100-240V ECG/NIBP/SPO2/Temp
Criticare 81H000XD
Item# Criticare-81H000XD
Regular price: $5,550.00
Sale price: $5,399.99

Product Description

 Monitor Patient VS nCompass <br>100-240V ECG/NIBP/SPO2/Temp <br>Criticare 81H000XD
Monitor Patient VS nCompass
100-240V ECG/NIBP/SPO2/Temp
Criticare 81H000XD

The nCompass™ 8100H Series monitor is the perfect choice for busy departments demanding accurate, reliable performance.

The 8100H Series monitor can monitor adult and pediatric patients in virtually every department. The innovative user interface features simplified menus and dedicated function keys for timely setups. The big, bright display with large numerics can easily be seen from across the room.

The 8100H Series monitor mounts quickly and easily to walls and rollstands.

Features and Benefits

Easy-to-read 12.1" active color TFT display. Easy to learn, easy to use with simplified menus and dedicated function keys. Mini-Trends can display condensed tabular vital signs data to show recent monitoring activities. Cardio-Respirogram Trend Display. Optional invasive cardiac output interface capability. Compatible with Criticare's 8500 Series Agent Gas Analyzer for anesthetic monitoring.

Technical Specifications

ECG

Number of Leads: 3- or 5-Lead Lead View: User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead); I, II, or III (3-Lead) Gain Selection: x0.5, x1, x2, and x4 Frequency Response: Diagnostic: 0.05 to 100 Hz (-3dB); Monitor: 0.5 to 40 Hz (-3dB); ST: 0.05 to 40 Hz (-3dB) Electrosurgery Protection: Yes Defibrillator Protection: Yes Pacer Detection/Rejection: Yes ST Accuracy: ±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm ST Range: -6.0 to 6.0 mm

NIBP (Non-Invasive Blood Pressure)

Technique: Oscillometric measure upon inflation Measurement Cycle: < 40 sec. typical Auto. Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr Max. Allowable Cuff Pressure: Adult/Pediatric: 300 mmHg; Neonate: 150 mmHg Range: Adult/Pediatric: 30-300 mmHg; Neonate: 20-150 mmHg Resolution: 1 mmHg Transducer Accuracy: ±2 mmHg or ±2% or reading, whichever is greater, over full range STAT Mode: 5 min of consecutive readings Clinical Accuracy: Complies with AAMI SP-10:2002

Pulse Oximetry

Range: 1% to 99% Resolution: 1% Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%; < 50%: unspecified Method: Dual wavelength LED

Heart (Pulse) Rate

Source: User selectable: Smart, ECG, Pleth, NIBP, IBP Range: 20 to 300 bpm (ECG, Pleth); 30-240 bpm (NIBP); 30-300 bpm (IBP) Accuracy: ±1 bpm or ±1% (ECG) whichever is greater

Temperature

Channels: 2 Range, Accuracy: 68° to 113°F (20° to 45°C): ±0.1°C Display Resolution: ±0.1° Probe: YSI-400 and YSI-700 Series

Respiration

Rate: 6 to 150 bpm (ECG); 2 to 120 bpm (CO2) Resolution: 1 bpm Accuracy: ±2 bpm or ±2%, whichever is greater Source: User selectable: Smart, CO2, ECG

CO2

Units: mmHg; %; kPa; Torr. Method: Non-dispersive infrared; autocalibrating. N2O Compensation: Manual (On/Off). Range: 0–99 mmHg; 0–12.5%; 0–12.5 kPa; 0–99Torr. Resolution: 1 mmHg; 0.1%; 0.1 kPa; 1 Torr. Flow, Sample Rate: 200 ml/min. Accuracy: ±2 mmHg; ±0.3 vol. %; ±0.3 kPa; ±2 Torr or ±4% of reading, whichever is greater, at 200 ml/min & respiratory rate of ≤ 120 bpm. Rise Time: 170 ms (10–90%) at 200 ml/min.

IBP (Invasive Blood Pressure)

Channels: 1 or 3. Transducer Sites: Selectable: ART, PA, LV, RV, LA, RA, CVP, ICP. Pressure Range: -10 mmHg to 300 mmHg. Transducer: Freq. Response 0–20 Hz; Impedance >300 ohms; Sensitivity 5μV/V/mmHg. Accuracy: ±1 mmHg or ±1%, whichever is greater.

Alarms

Indication: Audible; Visual (EN 60601-1-8) Silence: Yes; 2 minutes or permanent.

Trends

Memory: 24 hours at 30 second intervals. Display: Tabular, Graphical.

Display

Screen: 12.1” active color touch screen Waveforms: 6, maximum Waveform Types: ECG Leads I, II, III, aVR, aVL, aVF, V; CO2; Respiration; Pleth; IBP; breath-by-breath CO2; ST Histograms; Mini-Trends; OXY-CRG Sweep Speed: 6.25, 12.5, 25, or 50 mm/sec

Controls

Keys: 10; membrane-activated. Rotary Knob: Push and rotate; 24 steps/turn.

System Outputs

COM ports: Two RS232-compatible DB9 ports; USB (for software downloads only). Nurse Call: Contact switch; 3.5 mm stereo phone jack, 14 mm. Defib/Sync BNC connector. Video Port: Serial VGA compatible. DC Out: 5 VDC output for optional wireless accessory. External Module Port: 8-PIN circular connector.

Printer (optional)

Type: Thermal line printer. Paper Speed: 12.5, 25 or 50 mm/sec.

Physical Properties

Weight: 12 lbs; 5.4 kg. Size: 12” (H) x 14.5” (W) x 11.5” (D); 30.5 cm (H) x 36.8 cm (W) x 29.2 cm (D)

Power Requirements

Voltage: 100 to 240 VAC; 50/60 Hz. Power Consumption: 35 W, typical. Battery: 1 sealed lead acid. Battery Life: 2 hours, typical (3 hours, typical for base unit). Recharge Time: 4 to 6 hours, typical.

Environmental

Operating Temperature: 32° to 104°F (0° to 40°C). Storage Temperature: 23° to 122°F (-5° to 50°C). Operating Humidity: 15% to 90%, noncondensing. Operating Altitude: -300 to 3,000 meters.

Classification

Medical Device: Class II Equipment (IIb EU). Electrical Protection: Class I Equipment. Degree of Protection: Type CF, Defibrillator-proof. Protection Against Ingress: IPX1. Safety: EN 60601-1.

Custom Configurations

Standard: ECG, Respiration, NIBP, HR, SPO2, Temperature, Arrhythmia and ST analysis, and Nurse Call interface connection. Optional: CO2, 1 or 3 channel IBP, printer.